NPD Mini PCR Hood NPD-232 | Product Details
CLEAN AIR EQUIPMENTNPD Mini PCR Hood - Laboratory Cabinet
NPD Mini PCR HoodUnable to load the product image. Check the WordPress media URL.
NPD ECO TECHNOLOGIES

NPD Mini PCR Hood

Model: NPD-232 · PCR-2.5 | DNA Carryover Containment & PCR Setup Workstation

The NPD-232 Mini PCR Hood is a compact, partially recirculating vertical-downflow workstation for pre-PCR preparation. HEPA H14-filtered airflow and UV decontamination help protect molecular-biology workflows against airborne particles and DNA carryover contamination.

99.995%H14 HEPA at 0.1–0.2 µm MPPS
70% / 30%Recirculation / filtered exhaust
745 × 435Work area width × depth (mm)
1,200 luxWork area illumination
Configuration
LCD Digital ControllerBenchtop UnitOptional Stand
Configurations and optional accessories subject to your laboratory requirements.
HEPA H14High-efficiency particle filtration
Dedicated Pre-PCRDNA carryover control
LCD ControllerSoft-touch digital operation
AISI 304Removable stainless-steel worktable
Product Overview

Compact Protection for Sensitive PCR Preparation

Previously amplified DNA can contaminate new PCR reactions through airborne particles or residue on laboratory surfaces. The NPD Mini PCR Hood creates a controlled preparation area with HEPA-filtered vertical airflow, an inward air barrier and UV-based decontamination to help minimise unwanted carryover during PCR master-mix and reaction assembly.

Its compact benchtop footprint, glass-fronted work zone, LCD controller and removable stainless-steel worktable make it practical for molecular diagnostics, genomics and academic or research laboratories.

Product Advantages

Key Features & Benefits

Designed for clean, repeatable pre-PCR workflows.

01

HEPA H14 filtration

Main filter provides 99.995% collection efficiency at 0.1–0.2 µm MPPS.

02

DNA carryover control

UV decontamination and clean vertical downflow support contamination-conscious pre-PCR preparation.

03

Controlled airflow

Approximately 70% recirculation via main HEPA filter and 30% filtered exhaust.

04

LCD controller

Microprocessor-based digital LCD with soft-touch blower, lighting and UV controls.

05

SS 304 worktable

Removable, perforated AISI 304 stainless-steel work surface for easy cleaning.

06

Compact benchtop design

Space-saving work zone with hinged tempered-glass front window and optional stand.

Technical Information

Technical Specifications

Model NPD-232 / PCR-2.5, based on the supplied product PDF.

ParameterSpecification
MakeNPD
ModelNPD-232
Model codePCR-2.5
ApplicationPCR setup, DNA carryover containment, molecular-biology sample preparation
Work-chamber dimensions (W × D × H)745 × 435 × 500 mm
Overall dimensions (W × D × H)815 × 615 × 931 mm
Front-opening height180 mm
Airflow designPartially recirculating vertical downflow
Main filterHEPA H14
HEPA efficiency99.995% at 0.1–0.2 µm MPPS
Air recirculation70% through main HEPA H14 filter
Exhaust30% via filter with 99% gravimetric efficiency
WorktableRemovable perforated AISI 304 stainless steel
Front panelsTempered glass
ControllerMicroprocessor-based digital LCD
ControlsSoft-touch keys for fan, fluorescent lamp and UV lamp
BlowerSingle-speed centrifugal motor blower
Lighting intensity1,200 lux
Noise level≤65 dB(A) at 1.5 m, open field
Power supply220/240 V AC, 50 Hz
Power consumption200 W
Optional standPowder-coated MS tubular stand

Specifications may be customised for laboratory requirements. The PDF does not specify the UV lamp wavelength or wattage, so those values are intentionally omitted.

Operating Principle

How the Mini PCR Hood Works

Four stages for contamination-conscious PCR setup.

01

Prepare the work area

Clean the removable worktop and set up a dedicated pre-PCR workspace.

02

Start filtered airflow

The blower circulates air through the main HEPA H14 filter, supplying vertical downflow.

03

Prepare PCR reactions

Carry out master-mix assembly and reagent handling within the controlled work zone.

04

Decontaminate between runs

Use established DNA-cleaning procedures and UV only when the work area is unoccupied.

Filtration & Airflow

HEPA-Filtered Vertical Downflow

Room air enters through the front opening, creating an inward air barrier. The blower delivers filtered vertical downflow across the worktable. Approximately 70% of air recirculates through the main HEPA H14 filter while 30% passes through a separate exhaust filter rated at 99% gravimetric efficiency.

Main HEPA H14 filtration is specified at 99.995% collection efficiency at 0.1–0.2 µm MPPS.

DNA Carryover & Controls

Simple, Reliable Laboratory Operation

The LCD digital controller provides soft-touch controls for the centrifugal blower, fluorescent work light and UV lamp. UV exposure can help reduce residual DNA contamination between work sessions, alongside approved physical cleaning and DNA-decontamination procedures.

Tempered-glass panels, a hinged window and 1,200 lux task illumination support visibility and accurate sample handling.

Where It Is Used

Applications & Laboratories

✓PCR master-mix preparation
✓DNA and RNA sample handling
✓Molecular diagnostics & genomics
✓Pre-PCR reaction assembly
✓Clinical & research laboratories
✓Academic molecular-biology labs
Important application note: This PCR setup hood is for non-hazardous work and DNA carryover control. It is not a substitute for a Class II Biological Safety Cabinet when handling infectious agents, nor for a suitable chemical fume hood for hazardous vapours. UV should be operated only when the work area is unoccupied.
Frequently Asked Questions

Mini PCR Hood FAQs

What is the NPD Mini PCR Hood used for?

It provides a dedicated, contamination-controlled area for PCR setup, master-mix preparation and handling molecular-biology samples.

Which filter does the NPD-232 use?

The main filter is HEPA H14, specified at 99.995% collection efficiency at 0.1–0.2 µm MPPS. The exhaust filter is separately specified at 99% gravimetric efficiency.

What are its internal dimensions?

The work chamber measures 745 × 435 × 500 mm (W × D × H). Overall dimensions are 815 × 615 × 931 mm.

Does it include UV decontamination?

Yes. It includes UV-lamp control, but the supplied PDF does not give the UV wavelength or lamp wattage.

Can it replace a biological safety cabinet?

No. It is intended for non-hazardous pre-PCR preparation and is not a substitute for a Class II Biological Safety Cabinet or a chemical fume hood.

Is a stand included?

A powder-coated MS tubular stand is listed as an optional accessory.

NPD ECO TECHNOLOGIES

Need a Compact Pre-PCR Workstation?

Discuss the NPD-232 Mini PCR Hood, installation requirements and available accessories with our team.

Get Product Quote ↗